Key recommendation

   

 

 

 

 

 

 

 

 

 

 

TMC-7100 Marine CEMS


 

Application

 

With the global strict control of ship emission pollution, the International Maritime Organization (IMO) constantly puts forward and upgrades the emission restriction requirements to reduce the emissions of sulfide, nitrogen oxide, carbon dioxide and other pollutants. From January 1, 2020, the emission restriction order of version 0.5 in Annex VI of the International Convention for the prevention of ship pollution has been formally implemented. Except for the restricted waters, the allowable emission limit of sulfur compounds from ships sailing in global waters has been reduced from 3.5% of fuel sulfur to 0.5%. Therefore, some ships have installed exhaust gas cleaning system (EGCS) to meet the requirements of the Convention. The flue gas treated by EGCS must be installed with online continuous exhaust monitoring system (CEMS) to meet the requirements of IMO MEPC259 (68) guidelines and ensure the monitoring, recording and archiving of pollutant emission concentration.

 

TMC-7100 General Introduction

 

TMC-7100 measures the gas concentration upstream or downstream of the scrubber and transfers them to the vessel’s main control system. TMC-7100 system integrates Shimadzu URA209 NDIR analyzer, adopts internationally famous brand sampling probe, heated pipeline, condensers, sampling pump and marine air conditioner. Tailor-made modular design for the unique application of marine ships, is easy to install and maintain, and is suitable for long-term high temperature, high humidity and strong vibration environment. Meet the requirements of MEPC.259(68).

TMC-7100 consists of the following parts:

  • Sampling probe with auto back purging

  • Heated sample gas pipeline

  • EC conditioning system

  • Shimadzu NDIR analyzer

  • Integrated in painted SS cabinet with air conditioner

  

TMC-7100 Solution Benefit

 

Continuous measurement of SO2, CO2 (SO2/CO2 ratio) with proven URA209 NDIR analyzer, ensure the accuracy of SO2 test under the concentration range of 40 µmol/mol reliable

  • Can expand NOx and O2 monitoring module

  • Low maintenance due to auto-calibration and auto back purging

  • MARPOL referenced extractive gas sampling instruments

  • Friendly HMI, long-term & massive data storage, GPS positioning

  • RS485 communication, seamless connection with EGCS from different manufactures

  


TMC-7100 Core: Shimadzu URA209 NDIR Analyzer


High selective

  • Low Detection Limit

  • High measurement accuracy

  • Automatic calibration

  • Automatic SO2& CO2 test range switching

  • Flexible interface for remote maintenance

  • Economic

  • Stable


 


Heated   Probe


Certified by DNV-GL Classification   Society: No. TAA00002FW

Heated internal surface filtration, keep   probe chamber free of dust

One-hand 90° rotation to replace the built-in   filter, pops out without tools

Increased protection degree IP66 for easy   commissioning and maintenance

Particularly suitable for use in   high-vibration marine applications

115/230 V universal power supply

All   connections are fully insulated, no cold spots

 

 

 

  Cooler


Low wash out effects for SO2,   NO2 etc. Complies with EN 15267-3

 

Pluggable heat exchanger with high   thermal conductivity

Two-stage condensation to fully remove H2O

Unique constant cooling control system, dew   point stability 0.1 °C

Integrated cooling temperature analog   output, remote monitoring


 


Air   Conditioner

 

Widely used for Marine application   globally

 

Energy-efficient Blue e wall-mounted   cooling units with integral e-Comfort controller and RiNano coating on the   condenser as standard

External circuit IP 34; Internal circuit   IP 54

Operation (environment):   +10 °C...+55 °C; Setting range: +30 °C...+35 °C

0.85 kW, 230 V, 1~, 50/60 Hz, Sheet   steel, WHD: 280 x 550 x 280 mm

Refrigerant: R134a; Quantity: 280 g;   Global Warming Potential (GWP): 1430; CO₂   equivalent (CO₂e): 0.4 t


 

 

 

 

DNV GL Certificate                                    



   

 

 LR Certificate 

 


BV   Certificate 

 ABS   Certificate

 NK Certificate

 



 


 

TMC-7100 Specification


 

Measuring Principle

NDIR

Measuring Components

SO2,CO2

Measuring Range

SO2:0-50/500ppm

CO2:0-5%/15%

Response Time

4~7S

Accuracy

0.5%  F.S.

Zero Drift

<1% F.S.

Span Drift

<2% F.S.

Linearity Error

<1% F.S.

Detection Limit

<1% F.S.

Repeatability

<1% F.S.

Auto-Calibration

Yes

Sample Temperature

190℃

Sample Dust Load

Max 1g/nm3

Heated Pipeline Temperature

120℃~180℃

Cooler Temperature

2~5℃(Control   Accuracy ±0.1%)

Pump Capacity

Max 400L/h

Sample Gas Inlet

1.0~1.5L/min

Filter Accuracy

0.1um

Display

LCD

Operation Temperature (Analyzer)

-5~55℃

Power

220VAC, 2,000W~3,600W

Weight

~220Kg

Cabinet Dimension

1,350x460x600mm

Storage & Transportation Temperature

-20℃~60℃