Rosemount process products


 Rosemount pressure transmitter technical solution

 

Pressure transmitter leading brand, has HART agreement, capacitive principle and other invention patents. In China, nearly 300,000 units are installed on customer sites each year.


Our plan:

(1) The transmitter communication protocol adopts 4-20 mA HART communication, which is convenient for on-site debugging and configuration;

(2) The transmitter housing has a special dense coating, which effectively resists salt spray corrosion.


Our characteristics:

(1) For small pressure measurement, we adopt the patented, tried and tested capacitive principle.


The conventional resistive principle requires the pressure to be large enough to produce the change of the rheostatic value, so it is less sensitive to the measurement of small pressure. Its advantage is that it can withstand large pressure, such as 50Mpa, and its cost is economical.

 

The capacitive principle is naturally sensitive to small pressure changes, the disadvantage is high cost, can not measure too high pressure (more than 13.89mpa occasions). Pressure measurement below 5Kpa is often the measurement of flue gas pressure, and the accuracy of these pressure measurements directly affects the desulfurization efficiency.


Therefore, we adopted the tested capacitive measurement principle patented by rosemount for the pressure measurement below 5Kpa, and the pressure measurement below 5Kpa is accurate and reliable with an accuracy of ± 0.04%.

 

For ordinary 0~400Kpa and 0~1000Kpa applications, we choose 2088 series (general resistance principle) with accuracy ±0.065%.



 

 

                

 

                                   

                   电容式测量原理                          电阻式测量原理

 


(2) excellent resistance to salt spray corrosion.

Salt fog test followed standard: ASTM B117

Main parameter: Salt solution 5%, 1000 hours

–Pictures take after 100, 200, 400, 600, 800 and 1000 hours


Other parameters reference as below table

            

Parameters

Specification

Rosemount Parameters

Temp of the chamber

35°C (+1.1/ -1.7°C)

35°C per ASTM   B117

pH of the solution

6.5 - 7.2

not measured, salinity is monitored to be   4-6% (refract meter)

Ratio of salt to water

5:95

same per ASTM B117

Volume of the fog

1.0 - 2.0ml/hour

@12 psi = 1.59mL/hr

(per B117)