Perma Pure GASS 6900 Pretreatment System

 

Taking Nafion tube as the core and combining with Baldwin series CEMS sample gas treatment technology, bepure has developed a series of CEMS sample gas treatment system - GASS system, of which GASS 6900 sample gas treatment system is its typical representative.

                               



The GASS 6900 sealed nema-4x housing contains two temperature control zones. The first temperature control area (see the high temperature area marked red in FIG. 3) is controlled at 90 ~ 95℃ to ensure no condensation of flue gas in this area. Once filtered high temperature flue gas (with a primary filter in the high temperature sampling probe) enters the condensed particulate filter (secondary filter) to reduce the particulate impurities to 0.1 m. If there is acid mist, it can also condense here and be eliminated automatically.
   

 

The maximum control temperature of Nafion® dryer is 95℃, which ensures that no condensation water precipitates out above the dew point temperature of the flue gas. In the second temperature control zone (the normal temperature zone in FIG. 2), the sample gas passes through the remaining part of the dryer to further reduce the dew point. It is suggested that the flue gas dew point after the control treatment should be 0 ~ -5℃. In addition, you can choose from this section to install other accessories required for backblowing, such as a sampling pump or a purge dryer.
    

GASS 6900 is placed directly in position by the flue as close as possible to the sampling probe. The GASS 6900 is equipped with PLC system to control the filter temperature and blowback time, as well as the heating of the back panel in the high temperature area and the separate heating of Nafion tube in the high temperature area. After the flue gas is extracted from the flue, the pretreatment of dust removal and dehumidification can be carried out immediately. While ensuring the integrity of the analysis, the heating pipeline and the condenser with large amount of maintenance can be omitted, and the calibration time and standard gas can be saved. Only 220V power supply and 0.6-0.8mpa oil-free and particulate free compressed air are needed. Easy installation and low maintenance.

 

Technical innovation evaluation of GASS 6900


Ø In view of many problems caused by precipitation of condensed water, Nafion drying tube with gaseous dehumidification was used to solve them.
Ø The ff-250 flocculation filter with filter diameter of 0.1um was adopted to solve the problem of SO3 acid fog and particulate matter.
Ø Nafion drying tube with gaseous dehumidification was also used to solve the loss of SO2 dissolved in condensation water.
Ø The dew point monitoring problem of acid flue gas after treatment was solved by Defender dew point display (including Nafion tube technology).


Ø For continuous and dry blowback gas required by Nafion tube, HD heat - free dryer was adopted to solve the problem.

 

 

The GASS 6900 system, installed in situ, can divide most of the water vapor in the low-temperature and high-humidity flue gas in the "ultra-low emission" flue gas in the gas phase, and the dew point of the flue gas can be ensured to be below 0℃(4,518mg/Nm3).


    

GASS 6900 dehumidifies in the gaseous state, which completely solves the problem of condensation water precipitation, and also completely avoids the problem of acid gas dissolving into the condensed water and producing acid solution, which leads to the corrosion of the system.